Load cells

TanenBase is not tied to one load cell. Which cell is under the hive is a runtime setting, picked on the setup page (Scale → Load cell) and stored in ZMS, so one firmware image serves every build of the frame.

The setting selects the two constants of the temperature correction (../../src/features/measurement/tempcomp.c): a gain k_c in mg/K and a thermal lag τ in seconds. It does not touch the zero offset or the scale factor — those come from the tare and calibrate steps and are per unit regardless of the profile.

Profiles

id Profile k_c τ Status
0 Generic / unknown 0 mg/K 1500 s No correction. The weight follows the daily temperature — ≈350 g p-p on the reference frame.
1 Bosche H40A-C3-0150 17734 mg/K 1500 s Measured, 40 h sweep + 144 h field validation → H40A-C3-0150/
2 Steinberg SBS-PF-150 0 mg/K 1500 s Not characterised — field run planned.
3 Custom from the stored record from the stored record Your own measured gain and lag, entered on the setup page.
4 Zemic L6E / L6E3 0 mg/K 1500 s Not characterised.
5 TAL220 / TAL220B 0 mg/K 1500 s Not characterised.
6 Flintec PC / SB 0 mg/K 1500 s Not characterised.

One entry per model, not per capacity: k_c is an absolute mg/K dominated by the mount, so a 150 kg and a 200 kg version of the same cell in the same frame land close enough that the variant is not worth a row.

A named-but-uncharacterised profile behaves exactly like generic — gain 0, no correction. The label still earns its place: the unit records which cell is bolted under the hive, and a later firmware can fill the constants in without the beekeeper touching anything, because config_get_lc() resolves non-custom profiles from this table on every read rather than from the stored copy.

The ids are the GATT contract (LC_PROFILE_* in ../../src/config/config.h) — append, never reorder. That is why LC_PROFILE_CUSTOM sits at 3, in the middle: it was there before the named cells were added. The setup page lists it last anyway; the dropdown order and the id order are independent. CONFIG_TANENBASE_LC_DEFAULT_PROFILE picks what a freshly flashed unit runs until the setup page stores something.

Why an uncharacterised cell ships with gain 0

k_c is a property of the cell plus its mount, not of the cell model. The H40A's datasheet temperature effect on zero (±0.014 %FS/10 K → 2.1 g/K) is 8× smaller than the 17.7 g/K the assembled frame actually shows: most of the drift is the mount, not the element.

So a plausible-looking number copied from a datasheet — or from a different cell of the same class — is not a conservative default. If it has the wrong sign or magnitude the correction adds error, silently, and it looks like the cell is bad. Uncorrected is the honest starting point; anything else has to be measured.

Characterising a cell from a production week

The reference sweep was a bench run with a separate logger. You do not need one: a station already reports raw weight and probe temperature, which is the whole input to the fit. This is the route planned for the SBS-PF-150.

  1. Put the cell on a profile with gain 0 — its own named profile, or generic. This matters: the fit needs the uncorrected weight. A run logged with a correction already active has to be inverted first (H40A-C3-0150/analysis/field_validation.py does that, but it is avoidable work).

  2. Put a dead weight on it and leave it alone. 20–25 kg of something that does not evaporate, gain moisture or get walked on. No colony: a hive's real mass change is exactly the signal you are trying to separate the drift from. Outdoors, in the frame it will actually live in — the mount is most of what you are measuring.

  3. Shorten the uplink interval for the week. At the 4 h heartbeat default a week gives ~42 points, and the daily cycle is what carries the information. Set Senden alle to 15–30 minutes on the setup page (Einstellungen) for the duration; 300–650 samples is the range the reference fit worked in. Put it back afterwards — that cadence costs battery (see ../POWER_BUDGET.md).

  4. Collect a full week including one clear hot day and one cold night. The fit is driven by the temperature span: the reference run got 17.5 K. Rain is not fatal but it is a confounder — wood and concrete absorb hundreds of grams and give it back over days (H40A report §9).

  5. Export the uplinks. Any sink works (TTN Storage Integration, BEEP, your own MQTT log). The decoder in ../../ttndecoder/ gives weight_kg / Gewicht in kg and t / TempOut in °C. Reshape to the semicolon CSV the analysis script expects:

    time;T_C;W_raw_kg
    2026-08-10T00:03:17Z;23.0;22.11
    
  6. Fit it. Point H40A-C3-0150/analysis/tempcomp_analysis.py at the CSV. It reports k, sweeps τ, and prints the residual σ with and without the correction. Uneven sampling is fine — the filter uses the real Δt between rows — but a long gap re-seeds nothing, so check for holes.

  7. Sanity-check before shipping it. k_c = -k. If σ does not improve by a clear multiple, or the sign differs from the reference cell's, something else moved during the week — do not enter that number. Hold-out validation (§3 of the H40A report) is the cheap way to tell a fit from a coincidence.

  8. Enter it as “own measured values” (g/K and minutes), re-tare, and watch the dead weight for a few more days before trusting it.

Adding a profile

  1. Measure. Constant known load, 40 h minimum, one full daily temperature cycle, logging raw weight + probe temperature — on the bench, or from a production week as above. Method, pitfalls and the pass criteria are in H40A-C3-0150/README.md §9.
  2. Fit. python H40A-C3-0150/analysis/tempcomp_analysis.py against your CSV gives k, τ and the residual σ. Sanity-check it on a hold-out window (§3) — a fit that only works on its own data is not a profile.
  3. Replay it on the firmware arithmetic before believing it: gcc -O2 -std=c99 -o host_test H40A-C3-0150/firmware/host_test.c H40A-C3-0150/firmware/tanen_tempcomp.c -lm. The integer implementation has to land on the Python model's numbers.
  4. Enter it as “own measured values” on the setup page (gain in g/K, lag in minutes) and let it run on a dead weight for a few days before trusting it.
  5. Only then add it to lc_profiles[] in ../../src/config/config.c, with a directory here holding the data and the report.

One caveat that saves a re-measurement: τ describes the lag between the temperature probe and the cell body, so it belongs to the frame and the probe placement, not to the cell. Every characterised profile so far carries the same 1500 s, and the optimum is broad — 15–40 min all land within 1 g. It is k_c that needs the work.

Changing the profile does not invalidate the tare

The correction is k_c · (T_eff − T_ref), which is zero at T_ref for any gain. T_ref is the temperature captured at the last tare, so switching profiles leaves the zero where it was — no re-tare needed, only a re-tare after mechanical changes.